Teledyne Advanced Pollution InstrumentationA Teledyne Technologies Comapny

Home    Company Profile    What's New    Contact Us   

Problem Report Forms

Model 400 - Warranty Repair Questionnaire

A PDF version of this form is available here.

Company:
Contact name:
Telephone:
Fax:
Email:
Serial #:
Software Version:
Describe the Failure Symptoms:
List All Warning Messages:
Test values:
ParameterObserved valueUnitsNominal Range
Rangeppb/ppm100 - 10,000
Stabilppb<1
O3 MEAS @ ZEROmV2500 - 4800
O3 REF @ ZEROmV2500 - 4800
O3 MEAS @ SPANmV2500 - 4800
O3 REF @ SPANmV2500 - 4800
PressureIn-Hg-A~1"<amb.
Sample Flowcm³/min800 ± 10%
O3 Gen.mV125 - 5000
O3 DrivemV0 - 5000
Sample TempºC1 - 5 > box temp
Photo Lamp TempºC52 ± 1
O3 Gen. TempºC48 ± 1
DCPSmV2500 ± 200
O3 Slope 1.0 ± 0.15
O3 Offsetppb-8 - +15
Time HR:MN:SC
O3 Dark ValuemV125 ± 50
O3 SPAN Valueppb50 - 1000
Is unit leak checked?Yes No
Will the unit allow you to SPAN (push the SPAN button)?Yes No
Will the unit allow you to ZERO (push the ZERO button)?Yes No
What are you using for a source of SPAN gas?
When was the last time your source was calibrated?
If possible, please include a portion of the strip chart pertaining to the problem. Circle pertinent data.



Home | Terms of Use
e-mail us: api-sales@teledyne.com

Copyright © 2009 Teledyne Technologies Incorporated. All rights reserved.