Problem Report Forms

Model 200AH - Warranty Repair Questionnaire

A PDF version of this form is available here.

Company:
Contact name:
Telephone:
Fax:
Email:
Serial #:
Software Version:
Describe the Failure Symptoms:
List All Warning Messages:
Test values:
ParameterObserved valueUnitsNominal Range
Range ppm 5 - 5,000
Stabilityppb0.0 - 0.2
Sample Flowcm³/min300 ± 50 (Std)
Ozone Flowcm³/min250 ± 15
PMT@ZeromV<100
Norm PMT@SpanmV0 - 4,000
PMT@SpanmV0 - 4,000
Span Conc.ppb 
AzeromV-10 - +50
HVPSV400 - 700 constant
DCPSmV2,500 ± 200
Rcell TempºC50 ± 1
Box Temp.ºCAmb ± 5
PMT Temp.ºC7 ± 1
BLOCK Temp.ºC 50 ± 1
CONV Temp.ºC700 ± 10
Rcell Press.In-Hg-A2 - 10 in-hg-a constant
Samp. Press.In-Hg-A~1" <Ambient
NO Slope 1.0 ± 0.3
NOx Slope 1.0 ± 0.3
NO Offset 0 ± 25
NOx Offset 0 ± 25
E TestV2,000 ± 200
O TestmV100 ± 20
Is unit leak checked?Yes No
Important note:
This analyzer has a sample flow bypass in it (check the pneumatic diagram in your manual). If your sample flow in the Rcell is blocked but the bypass flow is still open then you are going to have sample flow through the rear panel. Make sure that you measure the sample flow through the rear panel with the bypass in the pneumatic path and with the bypass REMOVED and capped up. If you don't cap the sample bypass, you are going to get bad readings for the true sample flow.