Model 101E - Warranty Repair Questionnaire

A PDF version of this form is available here.

Company: *
Contact name: *
Telephone: *
Fax:
Email: *
Serial #:
Software Version:
Describe the Failure Symptoms:
List All Warning Messages:
(Note:Depending on options installed, not all test parameters shown below will be available in your instrument.)
Test values:
ParameterObserved valueUnitsNominal Range
Rangeppb/ppm50 ppb - 20 ppm
H2S STBppb≤ 1 ppb with zero air
SAMP FLcm3/min 650 ± 10%
PMT signal with zero airmV-20 to 150
PMT signal at span gas conc mV
ppb/ppm
0-5000
0-20,000 ppb
NORM PMT at span gas concmV
ppb/ppm
0-5000
0-20,000 ppb
UV LampmV 2000 to 4000
STR. LGTppm≤ 100 ppb/zero air
DARK PMTmV -50 to 200
DARK LAMPmV-50 to 200
SLOPE 1.0 ± 0.3
H2S OFFSmV< 250
HVPSV500-900
ETESTmV2000 mV ± 1000
OTESTmV2000 mV ± 1000
RCELL TEMP°C50 ±1
BOX TEMP°CAmbient +˜ 5
PMT TEMP°C7°C ± 2°
IZS TEMP°C 50°C ±3
Has the analyzer been check for leaks?
Yes  No
For proper flows?
Yes No
If possible, please include a portion of the strip chart pertaining to the problem. Circle pertinent data.